| |
Typical Characteristics Of Specified Analysers |
The analyser shall be selected so that the relevant ASTM Test Method on
or with which the
measurement is based or correlated is the same one as that used in the
site laboratory,
measuring the same property. The Principal shall be consulted if this
cannot be complied
with.
Analysers that are intended for environmental/pollution control or
Continuous Emission
Monitoring System (CEMS) may require approval or have to conform to the
requirements
specified by the local environmental/pollution control authority.
Key characteristics that are important to the service are listed below.
Composition
Multiple-component analysis can be best carried out with PGC. Mass
spectrometers have
the advantage of fast response times and are sometimes advantageous for
selective trace
component analysis in gaseous products. A typical response time is 6 s
per component.
The calibration matrix of a mass spectrometer is often
adversely affected by components which
were not included in the initial calibration mixture or which were
either under- or over-represented
in the initial calibration mixture. Calibration results may therefore
have limited validity. Certain mixtures may be difficult to analyse with a mass
spectrometer. For PGC, the calibration mixture for background gas should be as
close as possible to the
material composition of the process sample. This is particularly so for
the measurement of trace
contaminants.
Infrared and ultraviolet spectroscopic techniques are typically suitable
for single-component
analysis of gaseous products, and have some limited application for
liquid products.
Infrared and ultraviolet spectroscopic techniques, more specifically
Fourier Transform
Infrared and Ultraviolet Diode Array techniques, are also suitable for
multi-component
analysis in a limited number of applications. Any cross-sensitivity to
components not of
principal measurement interest should always be checked when infrared
and ultraviolet
spectroscopic techniques are used.
Modern PGCs usually offer the advantage of calculating derived physical
properties such
as heating value, density, distillation properties, and even octane
numbers. However, their
accuracies have to be checked to ensure that they meet the specified
requirements.
Density / Relative density
Density meters should be of the vibrating type. The use of radioactive
types requires the
approval of the Principal.
Dissolved oxygen
Dissolved oxygen is normally measured with a polarographic cell. The
output value is
dependent on the temperature due to the change in permeability of the
membrane at
different temperatures. The cell should be used in accordance with the
Manufacturer’s
specifications, e.g., for minimum flowing conditions. The sensor should
be kept clean and
unclogged at all time.
Metal-electrode-type dissolved oxygen meters often have robust
self-cleaning features. The
output is dependent on oxidation-reduction potentials (REDOX) of the
water. Heavy metals
and free oil may impede their use.
Flammable gas detection (in ambient air)
Need Input - Email Us
Flashpoint analysers
Flashpoint analysers that detect the flash on explosion should be used. The response of flashpoint analysers using semi-conductors or
pellistor type detectors is product
(i.e. feedstock) dependent. It is also sensitive to the presence and
build up of olefinic material e.g., in
catalytic cracked sample. For process samples with a sulpur content > 1
%, pellistor type or flash point
analysers shall not be used.
HCl measurement in process gas
Analysers using the change of conductivity of a reagent should be used.
Alternatively,
analysers using the coloration of paper tape impregnated with reagent
could be considered
where lower accuracy is acceptable.
H2S measurement
For measurement of H2S in ambient air, For
measurement of low
concentration H2S in process gas, the lead acetate paper coloration
principle or PGC with
flame photometric detector should be used. For higher concentrations (%
range), UV
spectrophotometer or PGC may be used.
KVP measurement
Where KVP is measured in order to obtain a correlation with RVP, an RVP
analyser should
be used directly.
Moisture measurements in gaseous products
-
Dewpoint mirror techniques are absolute measurements. They are costly
and are
specifically suitable for higher moisture concentrations (0.05 % and
higher). The
technique shall not be used to measure water in natural gas, due to the
difficulty in
obtaining consistent results. High hydrocarbon concentrations may interfere. Typically, the
hydrocarbon dewpoint should
not be higher than 10 °C above the water dewpoint.
-
Karl-Fischer-type titration is an absolute measurement. Output shall
be expressed in
mg/m3.
-
Metal-oxide-type sensor measurements are relative measurements. The
output shall
be calibrated against a test gas mixture or against an absolute
measurement. The
output calibration is dependent on temperature and pressure. For
accurate
measurements the flow cell shall be kept at constant pressure and
temperature. The
characteristics of aluminium-oxide-type sensors are, in general, not
stable and regular
verification of the calibration factors is therefore required (a typical
requirement is once
per year, but this may vary depending on the application).
-
P2O5 (phosphorus pentoxide) sensors are theoretically absolute
measurements. However, they shall be treated as a relative-type measurement and
calibrated with a test mixture or against an absolute measurement. This analyser type
shall not be used in process streams containing double-bonded hydrocarbons or those rich
in hydrogen. The analyser is flow-sensitive.
-
LiCl (lithium chloride) type sensors are applicable for Relative
Humidity measurements. Their main application is in buildings, for use as a tool
for climatic conditioning.
-
Hygroscopically coated vibrating crystal-type moisture meters have
the best accuracy. This type of meter should not be used for prolonged measurements of
moisture concentrations in excess of 2000 cm3/m3
NOTE: At moisture concentrations in excess of 2000 cm3/m3, the
hygroscopic layer on the crystal
may be washed off.
-
Measurement by conductivity of a hygroscopic salt-glycerol solution
is fairly suitable for natural gas application as the sensor is relatively easy to rejuvenate,
although this requires specialist attention.
-
Silicon oxide type sensors are thermally stable and less hygroscopic
than aluminium oxide type sensors. When equipped with a feature to momentarily heat the
sensor, they burn off any hydrocarbon contaminants and ‘left over’ moisture from
high loads, thus giving faster response and recovery. A silicon oxide type sensor is
preferred to metal oxide and P2O5 sensors.
-
Fibre-optic type sensors provide an in situ means of measurement and
measure dew point from - 70 °C to 10 °C at pressures up to 250 bar and temperature
from - 30 °C to 95 °C. It is a relative humidity measurement and converts results to
ppmv, dew point.
Octane analyser
Cold flame front analysers and octane engine type analysers have been in
use. Near
InfraRed analysers are preferred where product specifications can be
satisfied.
Oil in/on water
For oil in/on water measurements the following applications should be
considered:
Care must be taken with respect to correlation and cross-checking with
laboratory
measurement. The intent of the measurement shall be made clear and well
documented.
Wherever feasible/practicable, the analyser selected shall have the same
working principle
as that of the corresponding laboratory instrument.
a) Oil on water, e.g., oil sheen in open channels - This is a
qualitative detection only.
Laser Diode Infra-red reflectance-type measurements are suitable.
b) Oil emulsified/dispersed in water. Infrared-type meters, e.g.,
measuring hydrocarbons dissolved in an extraction fluid such as Freon,
have a reduced response for naphthenic hydrocarbons and do not
respond to the aromatic hydrocarbon content of the oil. Ultraviolet-type
meters do not respond to the naphthenic and the paraffinic content of
the oil. Instead they rely on the co-presence of aromatics in oil.
Typical solubility rates for paraffinic and aromatic hydrocarbons are
about 20 mg/l and 80 mg/l
respectively.
c) Total Organic Carbon - TOC (2.5.21). Any oil or hydrocarbon entering
the measuring
cell will be detected. The UV oxidation type is preferred to the “hot
oxidation” type of
TOC analyser. TOC measurement is good for measuring carbon content from
carbon
compounds or elemental carbon in an aqueous solution. Any oil that is
insoluble in
aqueous solution may not be suitable. If sample does not contain any
inorganic carbon
component, the acidification part of the analyser system need not be
included. Higher
range is possible with dilution. This type of analyser requires
extensive maintenance,
especially the “hot oxidation” type. The high consumption of reagent
should be
considered. For low range measurement, the standard sample should not be
stored for
long periods as it is not stable.
Oxygen
a) Zirconium-oxide-type meters are suitable for measuring oxygen in
combustion flue
gas. These meters shall not be used for gases containing flammable gas,
or unburned
material, e.g., soot. The sensor, which is kept at a temperature of
typically 800 °C,
should be of an Ex-d construction in accordance with IEC 60079 by means
of a flame
arrestor. For very high flue gas temperature applications, a cooling
extension tube
shall be used. The start-up of fired equipment may pose conditions of
increased hazard due to the likelihood of unburned fuel being present in
a cold environment, e.g., upon misfiring. Therefore, the design requires
careful assessment of the risks involved. Equipment selection shall take
into
consideration the suitability of the sensor with respect to the gas
group and temperature class.
For example, the presence of a high concentration of hydrogen in fuel
gas requires the sensor to
have an electrical certification allowing it to operate in such an
environment, while the
temperature class shall be related to the auto-ignition temperature of
the fuel.
For prolonged measurement at low levels (below 0.3 % oxygen) in an inert
atmosphere, e.g., nitrogen, zirconium-oxide-type sensors ‘fall asleep’
and require
many hours before they are again able to respond to oxygen. For this
reason,
zirconium-oxide-type sensors should not be used under these
circumstances.
b) Para-magnetic-type oxygen meters are suitable for oxygen measurement
in a
background of high inert gas or hydrocarbon. These sensors are also
preferred for the
measurement of high temperature regeneration flue gas where a high
loading of
particulate matter very rapidly clogs the filter of the in-situ
zirconium-oxide-type cell. If
the sample is to be returned to the process stream, the reading will be
affected by any
variation of backpressure. Certain gas species present in process
mixtures may affect
the measurement due to their para-magnetic properties.
c) Polarographic (electrochemical) type sensors are suitable for trace
oxygen analyses,
e.g., in Liquefied Natural Gas, or for applications where a precision of
5 % relative, or
worse, is acceptable.
2.5.14 pH measurements
a) Electrode-type pH meters should be used only for pH measurements 4 <
pH < 10.
Electrode-type pH meters shall not be used for pH > 12.
NOTE: Below a pH of 4, precision is reduced, and above a pH of 10,
precision and electrode life are
reduced.
Dual junction reference electrodes should be considered for fouling
conditions. Severe
services may require additional (automated) features to maintain clean
electrodes, i.e.,
mechanical (ultrasonic, brushes, flushing) and/or chemical cleaning. Low
conductivity
fluids, e.g., demineralised water, require a liquid-earthing electrode.
The construction
of the reference electrode should ensure an unimpeded delivery of
electrolyte from the
reference electrode to the process. Pressure compensated constructions
or sleeve
(fritt) type reference electrodes may satisfy that condition.
The use of an electrode system which immobilises the reference
electrolyte in an
ionically conducting, non-porous, hard polymer prevents fouling and
electrode
poisoning. This is preferred for fouling service and samples containing
H2S.
For low water content processes, e.g., in chemical processes, and under
process
pressures of –1 bar to 15 bar and temperatures of 0 °C to 140 °C, a
differential pH
sensor not using potassium chloride reference electrode should be used
for pH
between 4 and 10.
b) Titrator-type analysers should be used for pH < 4 and pH > 10.
When a fast-response system is required, the installing of a titrator in
combination with
an electrode-type pH meter should be considered for improved response.
c) For a fast-response pH measuring system for pH > 12, a Flow Injection
Analysis
analyser may be considered.
NOTE: Systems b) and c) may require a skilled maintenance pool. The
reagent consumption rate is
relatively high. The possibility of frequent breakdown should be
considered in the process design.
Phenol in water
Colorimetric type water quality analysers should be used.
NOTE: PGC shall not be used for this purpose because phenol can be
adsorbed and desorbed on the column material, thus causing low output
values when concentrations are high and high output values when
concentrations are low.
Pour Point Measurement
For pour point measurement, the detection of the motion of a controlled
disturbance to the
surface of a sample in a measuring cell is used. An analyser type shall
be selected
according to the measurement intent. Analysers are available which
conform to different
ASTM test methods. Improved resolution is available with some analysers
that follow a
different ASTM method than that adopted for finished product quality
conformance.
RVP
Reid Vapour Pressure analysers should be used. Where oxygenated fuels
are tested and where the ‘dry’ method is specified, DVPE has to be
calculated and reported. An analyser selected should be able to measure
and report the RVP and DVPE.
Salt in Crude
This is a batch type analysis in which the salt is released in the
sample matrix by mixing it
with alcohol and xylene. The conductivity of the extract is then
measured in the analyser.
Stack gas composition analysers
The analysers selected shall satisfy the requirements of the local
Environmental Regulatory
Authority. In the absence of local requirements, preference shall be
given to analysers that
meet the requirements of Authorities such as the EPA and TÜV. The
selection of the
analysers is generally dependent on whether the measurement is required
to be done on
the wet or dry basis. Typically, the following measurements can be
performed by stack gas composition analysers (in order of most common
measurements):
- Oxygen (ZrO2-type oxygen meter) by:
· in-situ type measurements, or
· extractive with analyser mounted against duct/stack wall.
- Opacity
· across stack with visible light optical type instrument.
- SO2, NOx
· across stack with infrared or UV light type instrument
· extractive with pre-dilution and chemo-luminescence type analysers
· extractive with infrared or UV light type instrument.
- CO
· across stack with infrared or UV light type instrument
· extractive with infrared or UV light type instrument.
- Dust/Particulates
· across stack with visible light optical type instrument.
Electronic equipment, other than that forming part of an in-situ
assembly associated with
stack measurements, should be located at ground level.
A complete Continuous Emission Monitoring System (CEMS) typically
consists of a few of
the above memtioned analysers together with a central data acquisition
system.
NOTE: The analysers shall be selected together with the sample take-off
and sample conditioning system as
a total package by the selected analyser Vendor.
Sulphur
For measurement of sulphur in distillate fuels down to 25 ppm, an x-ray
flourescence
principle should be used. For measurement of sulphur below 25 ppm, an
ultraviolet
flourescence technique shall be used. The use of PGC with FPD technique
for measurement of sulphur below 25 ppm
TOC in water
A distinction shall be made between the measurement of TOC (Total
Organic Carbon) to
meet environmental regulatory requirements and the measurement to detect
oil leakage
into steam condensate and/or steam system. For environmental
considerations, the
requirements are to be followed as specified. For measurement
considerations, the
technique selected shall be determined by the sensitivity and range of
measurement
required. TOC should be measured by the UV oxidation technique as a
first choice. The ‘hot catalytic
type’ shall be selected if the UV oxidation technique cannot assure the
completeness of
carbon detected and the sensitivity required. Because of the high
combustion temperature and the relatively slow cool-down of the oven
section, hot catalytic type analysers shall not be used in hazardous
areas, and should not be used in analyser houses (even if they are
ventilated).
The typical usable range for TOC measurements is 0 mg/l to 50 mg/l or
less. The lowest
practicable reading is limited to the background signal, which could be
typically 2 mg/l for
tap water, or even higher in systems where corrosion inhibitors are
applied.
NOTE: Compounds that have reduced oxidation efficiency with the UV
oxidation technique do not normally occur in petrochemical processes.
They would normally occur in minor concentrations compared to the bulk
of hydrocarbons. Typical reduced efficiencies are in general better than
80 %. The overall error in measuring organic carbon is normally better
than 2 % of the measured value.
The use of a variant of the PGC approach, stripping the hydrocarbons and
detecting them
by means of FID, could be considered for the ppb range of measurement of
light
hydrocarbon (Carbon < 12) contamination in water or steam samples. It
shall not be used
for residual fuel contaminated samples.
Viscosity
For optimum precision, capillary-type viscometers shall be used.
Other measuring principles, e.g., the tuning fork principle, may be
considered for
measurements on binary mixtures, or for applications requiring only a
rough precision (5 %
of the range or worse) such as in assessing the pumpability of heavy
fuel oil.
Water Analysis (Other properties)
Chlorine in fresh water or seawater
Residual or free chlorine should be stated. Different reagents are used
for each species of
chlorine. The same polarographic principle of releasing ions from
reagent reaction with
chlorine and measuring the current by electrodes in a flow cell should
be used.
Ammonium ion in sour water
Instruments of the on-line filtration type should be used. Cross
sensitivity to interferences
should be checked with vendors.
Alkalinity (m and p alkalinity)
On-line titration type analysers should be used.
COD analyser
On-line titration type analysers should be used.
Links
|